Precise efficiency calibration of an HPGe detector: source measurements and Monte Carlo calculations with sub-percent precision.

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine

J C Hardy, V E Iacob, M Sanchez-Vega, R T Effinger, P Lipnik, V E Mayes, D K Willis, R G Helmer

Affiliations

  1. Cyclotron Institute, Texas A&M University, College Station 77843-3366, USA. [email protected]

PMID: 11839060 DOI: 10.1016/s0969-8043(01)00168-3

Abstract

With the goal of measuring precise gamma-ray intensities for short-lived (< 5 s) accelerator-produced activities, we have calibrated the efficiency of an HPGe detector between 53 and 1836keV to sub-percent precision with a combination of source measurements and Monte Carlo calculations. Using known or independently measured detector dimensions, we have achieved both relative and absolute agreement (the latter, to 0.1%) between the calculated and measured efficiencies with only two adjustable detector parameters, the thicknesses of the contact dead layers.