A microwave plasma cavity assembly for atomic emission spectrometry.

Analytical and bioanalytical chemistry

H Matusiewicz

Affiliations

  1. Department of Analytical Chemistry, Politechnika Pozna?ska, 60-965, Pozna?, Poland.

PMID: 15045327 DOI: 10.1007/s0021663550623

Abstract

Although inductively coupled plasmas (ICPs) are widely used for multielement analysis microwave induced plasma (MIP) offers a great potential for a variety of applications. Modifications to incorporate MIP into commercial ICP direct reading spectrometer systems have been developed. A direct reading échelle spectrometer is described which opens new possibilities for the successful construction of commercial MIP-AES systems with the potential to run all of the typical methods worked out for earlier ICP-AES applications. Use of flow injection techniques and automation to couple with in situ concentration will likely offer a further improvement in the analytical performance of this system. Due to the capabilities demonstrated by this spectrometer it appears that hybrid instruments will be increasingly important for future developments in optical spectrometry. This is particularly true for very demanding areas such as atomic emission spectrometry. The system could be readily commercialized.