Anal Bioanal Chem. 1996 Jun;355(5):716-8. doi: 10.1007/s0021663550716.

Speciation of chromium by coupled column HPLC-ICP-MS - the effects of interfering ions.

Analytical and bioanalytical chemistry

M Pantsar-Kallio, P K Manninen

Affiliations

  1. Lahti Control and Research Laboratory, Niemenkatu 73 C, FIN-15210, Lahti, Finland.

PMID: 15045351 DOI: 10.1007/s0021663550716

Abstract

ICP-MS is used as an element selective detector for chromium species in aquatic samples using a coupled columm system consisting of a cation guard column and an anion column. The effects of large concentrations of disturbing ions, like NO(3)(-), Cl(-), CO(3)(2-) and SO(4)(2-), are discussed and the comparison with the use of different chromium isotopes for data acquisition is given. Quite low concentrations such as 100 mg/l of CO(3)(2-), 10 mg/l Cl(-) and even 5 mg/l SO(4)(2-) can cause an overestimation of chromium if not properly resolved1995 / Accepted: 24 October 1995