[Erroneous X-ray dose measurements in paediatric cardiac investigations].

RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin

[Article in German]
D G Onnasch, F Schröder, G Fischer, H-H Kramer

Affiliations

  1. Klinik für Kinderkardiologie, Christian-Albrechts-Universität zu Kiel. [email protected]

PMID: 15173980 DOI: 10.1055/s-2004-812930

Abstract

According to national and international rules the x-ray exposure of cardiac catheter examinations have to be measured. For a Philips Integris angiographic system used for paediatric heart catheterization we found severe errors in the indicated time of fluoroscopy and total dose area product ( DAP) both for fluoroscopy and radiography. Fluoroscopy times shorter than 6 seconds and DAPs smaller than 0.1 Gy cm (2) are ignored and not considered in the indicated total fluoroscopy time and the fluoroscopic and radiographic DAPs. This error leads to a systematic underestimation of the risk of radiation-induced cancer especially with infants and children.