Rev Sci Instrum. 2008 Aug;79(8):083110. doi: 10.1063/1.2972151.

A bent electrostatic ion beam trap for simultaneous measurements of fragmentation and ionization of cluster ions.

The Review of scientific instruments

O Aviv, Y Toker, M Errit, K G Bhushan, H B Pedersen, M L Rappaport, O Heber, D Schwalm, D Zajfman

Affiliations

  1. Department of Particle Physics, Weizmann Institute of Science, Rehovot 76100, Israel.

PMID: 19044339 DOI: 10.1063/1.2972151

Abstract

We describe a bent electrostatic ion beam trap in which cluster ions of several keV kinetic energy can be stored on a V-shaped trajectory by means of an electrostatic deflector placed between two electrostatic mirrors. While maintaining all the advantages of its linear counterpart [Zajfman et al., Phys. Rev. A 55, R1577 (1997); Dahan et al., Rev. Sci. Instrum. 69, 76 (1998)], such as long storage times, straight segments, and a field-free region for merged or crossed beam experiments, the bent trap allows for simultaneous measurement of charged and neutral fragments and determination of the average kinetic energy released in the fragmentation. These unique properties of the bent trap are illustrated by first results concerning the competition between delayed fragmentation and ionization of Al(n) (-) clusters after irradiation by a short laser pulse.