IEEE Trans Med Imaging. 1986;5(4):222-8. doi: 10.1109/TMI.1986.4307780.

Time-coded aperture tomography: experimental results.

IEEE transactions on medical imaging

J W Van Giessen, M A Viergever, C N De Graaf, H J Dane

PMID: 18244014 DOI: 10.1109/TMI.1986.4307780

Abstract

In this paper, we discuss the applicability of a time-coded aperture system especially designed for thyroid tomography on the basis of phantom experiments. Our studies show that 1) the quality of the reconstructions is high (e.g., a cold spot of 6 mm diameter in a thyroid phantom can easily be detected), and 2) the reconstruction can be carried out in less than 11 min on a standard 16 bit minicomputer (HP1000). It is therefore concluded that the clinical potentiality of the device is good.

Publication Types