IEEE Trans Appl Supercond. 2009 Jun;19(3):2281-2284. doi: 10.1109/TASC.2009.2018808.

An Analytical Technique to Elucidate Field Impurities From Manufacturing Uncertainties of an Double Pancake Type HTS Insert for High Field LTS/HTS NMR Magnets.

IEEE transactions on applied superconductivity : a publication of the IEEE Superconductivity Committee

Seung-Yong Hahn, Min Cheol Ahn, Emanuel Saul Bobrov, Juan Bascuñán, Yukikazu Iwasa

Affiliations

  1. Francis Bitter Magnet Laboratory, Massachusetts Institute of Technology, Cambridge, MA 02139 USA.

PMID: 20407595 PMCID: PMC2856088 DOI: 10.1109/TASC.2009.2018808

Abstract

This paper addresses adverse effects of dimensional uncertainties of an HTS insert assembled with double-pancake coils on spatial field homogeneity. Each DP coil was wound with Bi2223 tapes having dimensional tolerances larger than one order of magnitude of those accepted for LTS wires used in conventional NMR magnets. The paper presents: 1) dimensional variations measured in two LTS/HTS NMR magnets, 350 MHz (LH350) and 700 MHz (LH700), both built and operated at the Francis Bitter Magnet Laboratory; and 2) an analytical technique and its application to elucidate the field impurities measured with the two LTS/HTS magnets. Field impurities computed with the analytical model and those measured with the two LTS/HTS magnets agree quite well, demonstrating that this analytical technique is applicable to design a DP-assembled HTS insert with an improved field homogeneity for a high-field LTS/HTS NMR magnet.