Rev Sci Instrum. 2013 May;84(5):053304. doi: 10.1063/1.4803004.

High precision measurements of carbon disulfide negative ion mobility and diffusion.

The Review of scientific instruments

D P Snowden-Ifft, J-L Gauvreau

Affiliations

  1. Department of Physics, Occidental College, 1600 Campus Rd., Los Angeles, California 90041, USA.

PMID: 23742542 DOI: 10.1063/1.4803004

Abstract

High precision measurements were made of reduced mobility, lateral and longitudinal diffusion of CS2(-) ions in 40 Torr CS2 and a 30-10 Torr CS2-CF4 gas mixture. The reduced mobility was found to be 0.4645 ± 0.0006 cm(2) atm/s V in CS2 and 0.523 ± 0.001 cm(2) atm/s V in the CS2-CF4 gas mixture at STP. The lateral diffusion temperatures for these two gases (295 ± 15 K and 297 ± 6 K) were found to be in good agreement with room temperature. By contrast longitudinal diffusion temperature was found to be slightly elevated (319 ± 10 (stat) ± 8 (sys) K and 310 ± 20 (stat) ± 6 (sys) K) though given the errors, room temperature diffusion cannot be ruled out. For lateral diffusion, significant capture distances (0.21 ± 0.07 mm and 0.15 ± 0.03 mm) were measured while for longitudinal diffusion the results were not conclusive.

Publication Types