J Med Phys. 2013 Oct;38(4):185-8. doi: 10.4103/0971-6203.121196.

Radiation dose estimation of patients undergoing lumbar spine radiography.

Journal of medical physics

Prince Kwabena Gyekye, Adu Simon, Emi-Reynolds Geoffrey, Yeboah Johnson, Inkoom Stephen, Cynthia Kaikor Engmann, Wotorchi-Gordon Samuel

Affiliations

  1. Radiation Protection Institute, Ghana Atomic Energy Commission, Legon, Accra, Ghana.

PMID: 24672153 PMCID: PMC3958998 DOI: 10.4103/0971-6203.121196

Abstract

Radiation dose to organs of 100 adult patients undergoing lumbar spine (LS) radiography at a University Hospital have been assessed. Free in air kerma measurement using an ionization chamber was used for the patient dosimetry. Organ and effective dose to the patients were estimated using PCXMC (version 1.5) software. The organs that recorded significant dose due to LS radiography were lungs, stomach, liver, adrenals, kidney, pancreas, spleen, galbladder, and the heart. It was observed that the stomach recorded the highest dose (48.2 ± 1.2 μGy) for LS anteroposterior (AP). The spleen also recorded the highest dose (41.2 ± 0.5 μGy) for LS lateral (LAT). The mean entrance surface air kerma (ESAK) of LS LAT (122.2 μGy) was approximately twice that of LS AP (76.3 μGy), but the effective dose for both examinations were approximately the same (LS LAT = 8.6 μSv and LS AP = 10.4 μSv). The overall stochastic health effect of radiation to patients due to LS radiography in the University Hospital is independent of the projection of the examination (AP or LAT).

Keywords: Conventional X-ray; effective dose; lumbar spine; organ dose; radiography

References

  1. AJR Am J Roentgenol. 2008 Jun;190(6):1453-61 - PubMed
  2. Br J Radiol. 2009 Jan;82(973):1-12 - PubMed
  3. J ICRU. 2005 Dec;5(2):i - PubMed
  4. Phys Med Biol. 1998 Aug;43(8):2237-50 - PubMed

Publication Types