Abstract

Cantilever based sensors are a promising tool for a very diverse spectrum of biological sensors. They have been used for the detection of proteins, DNA, antigens, bacteria viruses and many other biologically relevant targets. Although cantilever sensing has been described for over 20 years, there are still no viable commercial cantilever-based sensing products on the market. Several reasons can be found for this - a lack of detailed understanding of the origin of signals being an important one. As a consequence application-relevant issues such as shelf life and robust protocols distinguishing targets from false responses have received very little attention. Here, we will discuss a cantilever sensing platform combined with an electrochemical system. The detected surface stress signal is modulated by applying a square wave potential to a gold coated cantilever. The square wave potential induces adsorption and desorption onto the gold electrode surface as well as possible structural changes of the target and probe molecules on the cantilever surface resulting in a measurable surface stress change. What sets this approach apart from regular cantilever sensing is that the quantification and identification of observed signals due to target-probe interactions are not only a function of stress value (i.e. amplitude), but also of the temporal evolution of the stress response as a function of the rate and magnitude of the applied potential change, and the limits of the potential change. This paper will discuss three issues that play an important role in future successful applications of cantilever-based sensing. First, we will discuss what is required to achieve a large surface stress signal to improve sensitivity. Second, a mechanism to achieve an optimal probe density is described that improves the signal-to-noise ratio and response times of the sensor. Lastly, lifetime and long term measurements are discussed.

Keywords: Biosensor; Cantilever sensing; Electrochemistry; Oligonucleotide; Surface stress

References

  1. Bioconjug Chem. 2008 Dec;19(12):2285-96 - PubMed
  2. Ultramicroscopy. 2007 Apr-May;107(4-5):422-30 - PubMed
  3. Anal Chem. 2007 Nov 1;79(21):8136-43 - PubMed
  4. Nano Lett. 2006 Apr;6(4):583-6 - PubMed
  5. Nat Nanotechnol. 2012 Sep;7(9):602-8 - PubMed
  6. Anal Chem. 1998 Nov 15;70(22):4670-7 - PubMed
  7. Phys Rev Lett. 2010 Dec 31;105(26):263903 - PubMed
  8. Nat Nanotechnol. 2009 Mar;4(3):179-85 - PubMed
  9. Langmuir. 2007 Jan 2;23(1):292-6 - PubMed
  10. Langmuir. 2013 Aug 13;29(32):9951-7 - PubMed
  11. Phys Rev E Stat Nonlin Soft Matter Phys. 2012 May;85(5 Pt 2):056313 - PubMed
  12. Nanotechnology. 2010 Feb 19;21(7):75501 - PubMed
  13. Analyst. 2008 Jul;133(7):855-63 - PubMed
  14. Nat Nanotechnol. 2008 Sep;3(9):533-7 - PubMed

Publication Types