Appl Radiat Isot. 2016 Mar;109:539-543. doi: 10.1016/j.apradiso.2015.11.036. Epub 2015 Nov 21.

Comparison of LabSOCS and GESPECOR codes used in gamma-ray spectrometry.

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine

L Done, L C Tugulan, D Gurau, F Dragolici, C Alexandru

Affiliations

  1. Horia Hulubei National Institute for Physics and Nuclear Engineering, IFIN-HH, POB MG-6, Magurele, Bucharest 077125, Romania; University of Bucharest, Doctoral School on Physics, POB MG-11, Magurele, Bucharest 077125, Romania. Electronic address: [email protected].
  2. Horia Hulubei National Institute for Physics and Nuclear Engineering, IFIN-HH, POB MG-6, Magurele, Bucharest 077125, Romania.

PMID: 26625727 DOI: 10.1016/j.apradiso.2015.11.036

Abstract

Two dedicated software packages -LabSOCS and GESPECOR- for efficiency evaluation in gamma-ray spectrometry, were compared for equivalence. The detection efficiency and the coincidence-summing corrections coefficients were calculated for a specific HPGe detector, for different sample parameters and energies typically encountered in environmental radioactivity measurements. The discrepancy between the results obtained with the two codes were acceptable for most of the applications. Furthermore, the deviations between the values of the standard sources/ reference materials activities from the certificate and the values obtained after Monte Carlo simulation were less than 8% for LabSOCS and 9% for GESPECOR.

Copyright © 2015 Elsevier Ltd. All rights reserved.

Keywords: GESPECOR; Gamma-ray spectrometry; LabSOCS; Monte Carlo; Volume sample