Electrical probe characteristic recovery by measuring only one time-dependent parameter.

The Review of scientific instruments

C Costin, G Popa, V Anita

Affiliations

  1. Iasi Plasma Advanced Research Center (IPARC), Faculty of Physics, Alexandru Ioan Cuza University of Iasi, Bd. Carol I nr. 11, 700506 Iasi, Romania.

PMID: 27036776 DOI: 10.1063/1.4943669

Abstract

Two straightforward methods for recovering the current-voltage characteristic of an electrical probe are proposed. Basically, they consist of replacing the usual power supply from the probe circuit with a capacitor which can be charged or discharged by the probe current drained from the plasma. The experiment requires the registration of only one time-dependent electrical parameter, either the probe current or the probe voltage. The corresponding time-dependence of the second parameter, the probe voltage, or the probe current, respectively, can be calculated using an integral or a differential relation and the current-voltage characteristic of the probe can be obtained.