Med Devices (Auckl). 2016 Mar 22;9:455-66. doi: 10.2147/MDER.S103102. eCollection 2016.
Clinical effectiveness and safety of powered exoskeleton-assisted walking in patients with spinal cord injury: systematic review with meta-analysis.
Medical devices (Auckland, N.Z.)
Larry E Miller, Angela K Zimmermann, William G Herbert
Affiliations
Affiliations
- Miller Scientific Consulting, Inc., Asheville, NC, USA.
- Miller Scientific Consulting, Inc., Asheville, NC, USA; Department of Human Nutrition, Foods, and Exercise, Virginia Tech, Blacksburg, VA, USA.
PMID: 27042146
PMCID: PMC4809334 DOI: 10.2147/MDER.S103102
Abstract
BACKGROUND: Powered exoskeletons are designed to safely facilitate ambulation in patients with spinal cord injury (SCI). We conducted the first meta-analysis of the available published research on the clinical effectiveness and safety of powered exoskeletons in SCI patients.
METHODS: MEDLINE and EMBASE databases were searched for studies of powered exoskeleton-assisted walking in patients with SCI. Main outcomes were analyzed using fixed and random effects meta-analysis models.
RESULTS: A total of 14 studies (eight ReWalk™, three Ekso™, two Indego(®), and one unspecified exoskeleton) representing 111 patients were included in the analysis. Training programs were typically conducted three times per week, 60-120 minutes per session, for 1-24 weeks. Ten studies utilized flat indoor surfaces for training and four studies incorporated complex training, including walking outdoors, navigating obstacles, climbing and descending stairs, and performing activities of daily living. Following the exoskeleton training program, 76% of patients were able to ambulate with no physical assistance. The weighted mean distance for the 6-minute walk test was 98 m. The physiologic demand of powered exoskeleton-assisted walking was 3.3 metabolic equivalents and rating of perceived exertion was 10 on the Borg 6-20 scale, comparable to self-reported exertion of an able-bodied person walking at 3 miles per hour. Improvements in spasticity and bowel movement regularity were reported in 38% and 61% of patients, respectively. No serious adverse events occurred. The incidence of fall at any time during training was 4.4%, all occurring while tethered using a first-generation exoskeleton and none resulting in injury. The incidence of bone fracture during training was 3.4%. These risks have since been mitigated with newer generation exoskeletons and refinements to patient eligibility criteria.
CONCLUSION: Powered exoskeletons allow patients with SCI to safely ambulate in real-world settings at a physical activity intensity conducive to prolonged use and known to yield health benefits.
Keywords: Ekso; Indego; ReWalk; SCI; ambulation; orthosis
References
- NeuroRehabilitation. 2015;37(3):321-40 - PubMed
- Disabil Rehabil Assist Technol. 2016 Oct;11(7):535-47 - PubMed
- JAMA Intern Med. 2015 Jun;175(6):959-67 - PubMed
- J Spinal Cord Med. 2012 Mar;35(2):96-101 - PubMed
- Biomed Res Int. 2014;2014:893468 - PubMed
- Spinal Cord. 2000 Sep;38(9):555-8 - PubMed
- J Spinal Cord Med. 2013 Jul;36(4):313-21 - PubMed
- World J Orthop. 2015 Jan 18;6(1):24-33 - PubMed
- Med Sci Sports Exerc. 1982;14(5):377-81 - PubMed
- Med Sci Sports Exerc. 2010 Apr;42(4):691-700 - PubMed
- Am J Phys Med Rehabil. 2012 Nov;91(11):911-21 - PubMed
- Top Spinal Cord Inj Rehabil. 2015 Spring;21(2):93-9 - PubMed
- Top Spinal Cord Inj Rehabil. 2015 Spring;21(2):100-9 - PubMed
- Top Spinal Cord Inj Rehabil. 2015 Spring;21(2):122-32 - PubMed
- Br Med J (Clin Res Ed). 1982 May 29;284(6329):1607-8 - PubMed
- Brain Res. 2015 Sep 4;1619:12-21 - PubMed
- Spinal Cord. 2013 Dec;51(12):882-92 - PubMed
- BMJ. 2003 Sep 6;327(7414):557-60 - PubMed
- Spinal Cord. 2014 Feb;52(2):110-6 - PubMed
- Arch Phys Med Rehabil. 1997 Jun;78(6):582-6 - PubMed
- Ann Intern Med. 2009 Aug 18;151(4):W65-94 - PubMed
- J Neuroeng Rehabil. 2015 Oct 14;12:82 - PubMed
- Top Spinal Cord Inj Rehabil. 2015 Spring;21(2):110-21 - PubMed
- Paraplegia. 1995 Jul;33(7):409-15 - PubMed
- Clin Rehabil. 2016 Jan;30(1):73-84 - PubMed
- Arch Phys Med Rehabil. 1995 Aug;76(8):779-83 - PubMed
- Arch Phys Med Rehabil. 2014 Oct;95(10):1878-1887.e4 - PubMed
- Int J Behav Nutr Phys Act. 2015 Sep 30;12:121 - PubMed
- J Rehabil Res Dev. 2015;52(2):147-58 - PubMed
- Spinal Cord. 2006 Aug;44(8):487-94 - PubMed
- Eur J Appl Physiol. 2014 Aug;114(8):1563-71 - PubMed
- Spinal Cord. 2013 May;51(5):356-9 - PubMed
- Spinal Cord. 2008 Jul;46(7):500-6 - PubMed
- Eur J Appl Physiol. 2013 Dec;113(12 ):3027-37 - PubMed
- Spinal Cord. 2011 Nov;49(11):1103-27 - PubMed
- IEEE Int Conf Rehabil Robot. 2013 Jun;2013:6650469 - PubMed
Publication Types