Abstract

Edge illumination x-ray phase contrast tomography is a recently developed imaging technique which enables three-dimensional visualisation of low-absorbing materials. Dedicated phase retrieval algorithms can provide separate computed tomography (CT) maps of sample absorption, refraction and scattering properties. In this paper we propose a novel "modified local retrieval" method which is capable of accurately retrieving sample properties in a range of realistic, non-ideal imaging environments. These include system misalignment, defects in the used optical elements and system geometry variations over time due to vibrations or temperature fluctuations. System instabilities were analysed, modelled and incorporated into a simulation study. As a result, an additional modification was introduced to the retrieval procedure to account for changes in the imaging system over time, as well as local variations over the field of view. The performance of the proposed method was evaluated in comparison to a previously used "global retrieval" method by applying both approaches to experimental CT data of a rat's heart acquired in a non-ideal environment. The use of the proposed method resulted in the removal of major artefacts, leading to a significant improvement in image quality. This method will therefore enable acquiring high-resolution, reliable CT data of large samples in realistic settings.

References

  1. Proc Natl Acad Sci U S A. 2012 Aug 28;109(35):13922-7 - PubMed
  2. Appl Opt. 2013 Oct 1;52(28):6940-7 - PubMed
  3. Philos Trans A Math Phys Eng Sci. 2014 Jan 27;372(2010):20130021 - PubMed
  4. Phys Rev Lett. 2013 Mar 29;110(13):138105 - PubMed
  5. Phys Med Biol. 2013 Jan 7;58(1):R1-35 - PubMed
  6. J Synchrotron Radiat. 2015 Jul;22(4):1072-7 - PubMed
  7. Rev Sci Instrum. 2011 Jul;82(7):073702 - PubMed
  8. Opt Express. 2011 Dec 5;19(25):25545-58 - PubMed
  9. Opt Express. 2014 Apr 7;22(7):7989-8000 - PubMed
  10. Opt Express. 2014 Jun 30;22(13):15514-29 - PubMed
  11. Med Phys. 2014 Jul;41(7):070701 - PubMed
  12. Phys Rev Lett. 2007 Mar 9;98(10):108105 - PubMed

Publication Types

Grant support