Pulse (Basel). 2018 Jul;6(1):9-18. doi: 10.1159/000484940. Epub 2018 Feb 15.

Evaluation of the Accuracy of a New Cuffless Magnetoplethysmography Blood Pressure Monitor in Hypertensive Patients.

Pulse (Basel, Switzerland)

Sua Kim, Jung Dong Lee, Jeong Bae Park, Seungjin Jang, Jungchae Kim, Sang-Suk Lee

Affiliations

  1. Department of Critical Care Medicine, Samsung Medical Center, Sungkyunkwan University, Seoul, Republic of Korea.
  2. JB Lab and Clinic, Seoul, Republic of Korea.
  3. Intelligence Lab, Convergence Center, LG Electronics, Seoul, Republic of Korea.
  4. Artificial Intelligence Lab, SW Center, LG Electronics, Seoul, Republic of Korea.
  5. Department of Oriental Biomedical Engineering, Sangji University, Wonju, Republic of Korea.

PMID: 30283749 PMCID: PMC6140613 DOI: 10.1159/000484940

Abstract

OBJECTIVE: There is a growing demand for cuffless blood pressure (BP) measurement as an easy alternative to cuff-occlusion-based BP measurement. We assessed the accuracy of a new cuffless, watch-style BP monitor with a magnetoplethysmography (MPG) sensor compared to two standard auscultatory and oscillatory BP monitors.

SUBJECTS AND METHODS: A total of 34 patients with uncontrolled hypertension (systolic BP ≥150 mm Hg or diastolic BP ≥95 mm Hg) were enrolled in the study. BP was measured by two conventional monitors and the new device during the pre-exercise phase, during isometric handgrip exercise, and during the recovery phase (5 min after exercise). The correlation between monitors was assessed using intraclass correlation coefficient (ICC) and Bland-Altman plots.

RESULTS: Although two reference monitors produced highly correlated BP measurements, each was differentially correlated with BP measurements obtained by the new MPG monitor. During exercise, the mean difference between systolic BP obtained by the MPG and oscillatory monitors was >7 mm Hg with an ICC of 0.549 (95$ CI 0.264-0.746) in systole and 0.737 (95$ CI 0.534-0.859) in diastole. The ICC between the auscultatory monitor and the MPG monitor was 0.753 (95$ CI 0.559-0.868) in systole and 0.841 (95$ CI 0.706-0.918) in diastole after exercise. Bland-Altman plots also indicated that the performance of the new MPG device was very similar to that of the auscultatory monitor.

CONCLUSION: Although the performance of the new MPG monitor was comparable to that of the reference monitors used in this study, improved stability and accuracy are necessary for accurate BP evaluation during dynamic activity.

Keywords: Blood pressure measurement; Cuffless device; Hypertension; Magnetoplethysmography sensor

References

  1. Curr Cardiol Rep. 2016 Nov;18(11):112 - PubMed
  2. Lancet. 1997 Sep 13;350(9080):757-64 - PubMed
  3. J Med Internet Res. 2016 May 05;18(5):e85 - PubMed
  4. JAMA. 1991 Jun 26;265(24):3255-64 - PubMed
  5. J Am Soc Hypertens. 2016 Dec;10(12):947-953.e5 - PubMed
  6. Curr Hypertens Rep. 2016 Jun;18(6):49 - PubMed
  7. PLoS One. 2016 Jun 01;11(6):e0155677 - PubMed
  8. J Hypertens. 2015 Aug;33(8):1580-7 - PubMed
  9. JAMA. 2013 Jul 3;310(1):46-56 - PubMed
  10. Zhongguo Yi Liao Qi Xie Za Zhi. 2015 Jul;39(4):285-7 - PubMed
  11. Adv Exp Med Biol. 2017;956:109-116 - PubMed
  12. Korean Circ J. 2016 Sep;46(5):681-687 - PubMed
  13. N Engl J Med. 2008 May 1;358(18):1887-98 - PubMed
  14. Hypertens Res. 2016 Aug;39(8):612-7 - PubMed
  15. Biometrics. 1977 Mar;33(1):159-74 - PubMed
  16. Blood Press Monit. 2016 Dec;21(6):340-344 - PubMed
  17. Am J Epidemiol. 2017 Feb 1;185(3):194-202 - PubMed
  18. J Am Heart Assoc. 2016 Dec 15;5(12): - PubMed

Publication Types