Robustness of aerosol delivery of amikacin liposome inhalation suspension using the eFlow® Technology.

European journal of pharmaceutics and biopharmaceutics : official journal of Arbeitsgemeinschaft fur Pharmazeutische Verfahrenstechnik e.V

Zhili Li, Walter Perkins, David Cipolla

Affiliations

  1. Insmed Incorporated, 700 US 206 North, Bridgewater, NJ 08807, USA.
  2. Insmed Incorporated, 700 US 206 North, Bridgewater, NJ 08807, USA. Electronic address: [email protected].

PMID: 34082122 DOI: 10.1016/j.ejpb.2021.05.021

Abstract

The purpose of these studies was to understand the effect on product performance of batch-to-batch variability in both the amikacin liposome inhalation suspension (ALIS) formulation and its delivery device, the Lamira® nebulizer system, designed and manufactured by PARI (PARI Pharma GmbH, Munich, Germany). Three batches of ALIS spanning a range of lipid concentrations (43, 48 and 54 mg/mL) were tested with nine PARI inhalation devices that varied within the production process of the vibrating membrane with respect to hole geometry. Three hole geometry clusters were built including a geometry close to the mean geometry (median) and two geometries deviating from the mean geometry with smaller (smaller) and larger (larger) holes. The output parameters included the nebulization rate, the aerosol droplet size distribution, the liposome vesicle size post-nebulization, and the fraction of amikacin that remained encapsulated post-nebulization. Across the 27 experimental combinations of three formulation batches and nine devices, the nebulization time varied between 12 and 15 min with the fastest nebulization rate occurring with the combination of low lipid concentration and larger hole geometry (0.68 g/min) and the slowest nebulization rate occurring with the combination of high lipid concentration and the smaller hole geometry (0.59 g/min). The mean liposome vesicle size post-nebulization ranged from 269 to 296 nm across all experimental combinations which was unchanged from the control samples (276-292 nm). While all three batches contained > 99% encapsulated amikacin prior to nebulization, the nebulization process resulted in a consistent generation of ~ 35% unencapsulated amikacin (range: 33.8% to 37.6%). There was no statistically significant difference in the generated aerosol particle size distributions. The mass median aerodynamic diameters (MMAD) ranged from 4.78 µm to 4.98 µm, the geometric standard deviations (GSD) ranged from 1.61 to 1.66, and the aerosol fine particle fraction (FPF < 5 µm) ranged from 50.3 to 53.5%. The emitted dose (ED) of amikacin ranged from 473 to 523 mg (80.2 to 89.3% of loaded dose (LD)) and the fine particle dose (FPD < 5 µm) ranged from 244 to 278 mg (41.4 to 47.1% of label claim (LC)). In conclusion, while variations in the lipid concentration of the ALIS formulation and the device hole geometry had a small but significant impact on nebulization time, the critical aerosol performance parameters were maintained and remained within acceptable limits.

Copyright © 2021 The Author(s). Published by Elsevier B.V. All rights reserved.

Keywords: ALIS; ARIKAYCE; Amikacin; Drug delivery; Encapsulation; Liposomes; Nebulized aerosol